• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIPRO CORPORATION (ODT) NIPRO ELISIO-H DIALYZERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIPRO CORPORATION (ODT) NIPRO ELISIO-H DIALYZERS Back to Search Results
Model Number DD+ELISIO-19H
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code Anemia (1706)
Type of Reportable Event Malfunction
Event or Problem Description
Clinic stated that they have experienced an increase in clotted dialyzers since switching from optiflux dialyzers to elisio dialyzers.Treatments are interrupted, and most are unable to return the patient's blood.The clotting is not patient, dialyzer size, or user specific.3 total incidents: (b)(6) 2021 - 1 incident (patient required blood transfusion on (b)(6) 2021) (b)(6) 2021 - 1 incident (b)(6) 2021 - 1 incident clinical specialist contacted the facility and offered in-person product training for may 3, 4, and 5th 2021.Clinic was advised to follow ifu for proper priming instructions, because the staff was following clinic protocols and not the product ifu.After our clinical specialist visited the clinic and observed the staff during priming and recirculating process, it was identified that the clinic was recirculating with 100ml instead of 200ml as instructed in the ifu.Arterial chambers were also observed not be be 75% full in some cases.During her visit to the clinic, the nurse manager advised that a patient required a blood transfusion 2 days after treatment was interrupted and the patient's blood could not be returned due to the clotted dialyzer.All data was documented and an email with her findings was sent to the nurse manager, providing instructions for proper priming, recirc, and levels of bloodline chambers.Initial date of occurence is unknown.Clinic stated the first incident occurred sometime in (b)(6).Dialzyer size was later received 04/28/2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIPRO ELISIO-H DIALYZERS
Common Device Name
DIALYZER
Manufacturer (Section D)
NIPRO CORPORATION (ODT)
8-7, hanuki-yachi, niida-aza
ohdate-shi, akita 018-5 794
JA  018-5794
MDR Report Key11899052
Report Number1056186-2021-00014
Device Sequence Number4403586
Product Code KDI
UDI-Device Identifier00383790007122
UDI-Public00383790007122
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/27/2021
Report Date (Section F) 05/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Model NumberDD+ELISIO-19H
Device Catalogue NumberDD+ELISIO-21H
Device Lot Number20J01A
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date04/01/2021
Device Age6 MO
Event Location Outpatient Treatment Facility
Date Report to Manufacturer05/27/2021
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date05/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Date Report Sent to FDA05/27/2021
-
-