| Model Number |
5541-A-402 |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Codes |
Pain (1994); Inadequate Osseointegration (2646)
|
| Date of Event |
03/13/2013
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
In performing follow-up, rep provided a revision operative report for (b)(6) 2013 for the patient's right knee.Preoperative diagnosis: loose right revision knee arthroplasty with maltracking of patella.Op report notes: "in less than 1 year she developed increasing pain and discomfort.Studies are suspicious for possible loosening of femoral component.There was only some distal cement around the femoral component.The stem appeared to be reactive line around it suggesting loosening of femoral component.The patella was also tilting and maltracking¿femoral component came out very easily with no fixation at all.The cement was bonded to the implant with augments.The stem came out and there was a fibrous membrane around the stem completely.There was no cement on the stem at all consistent with femoral component loosening.We had to remove the patellar component to position it more centrally.".
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding loosening involving a triathlon augment was reported.The event was confirmed via medical review.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: a review of the provided medical information by a clinical consultant indicated: there were revision surgeries for femoral loosening and patellar mal-tracking but the work-up and etiology for these were not clear.The last revision was result of a failure of the femoral component-stem construct at the threaded junction.The likely cause of the failure was micromotion at the threaded junction led to wear and material fatigue.To determine the mode of failure additional materials would be required as well as evaluation of the explant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: a review of the provided medical information by a clinical consultant indicated: there were revision surgeries for femoral loosening and patellar mal-tracking but the work-up and etiology for these were not clear.The last revision was result of a failure of the femoral component-stem construct at the threaded junction.The likely cause of the failure was micromotion at the threaded junction led to wear and material fatigue.To determine the mode of failure additional materials would be required as well as evaluation of the explant.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device is needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
In performing follow-up, rep provided a revision operative report for (b)(6) 2013 for the patient's right knee.Preoperative diagnosis: loose right revision knee arthroplasty with maltracking of patella.Op report notes: "in less than 1 year she developed increasing pain and discomfort.Studies are suspicious for possible loosening of femoral component.There was only some distal cement around the femoral component.The stem appeared to be reactive line around it suggesting loosening of femoral component.The patella was also tilting and maltracking¿femoral component came out very easily with no fixation at all.The cement was bonded to the implant with augments.The stem came out and there was a fibrous membrane around the stem completely.There was no cement on the stem at all consistent with femoral component loosening.We had to remove the patellar component to position it more centrally.".
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding loosening involving a triathlon augment was reported.The event was confirmed via medical review.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical information by a clinical consultant indicated: conclusion/assessment: the records were not entirely complete, and no dated or sequential radiographs were provided for review.The patient had a primary knee performed and just over 6 months later had an arthroscopy for the it band and patellar tilt.The records gapped until 2012, when after having pain, the patient had a revision for a loose femur.She did well initially, but 1 ½ years later was seen for pain and instability and ultimately underwent an arthroscopy after an aspiration showed metal stained fluid.She then underwent revision of the femur, and it was found that she had broken the femoral stem at the junction with the threads.It was commented that the failure could represent corrosion.The revision became infected and she underwent a wash out and polyethylene exchange 6 months later.No records were provided for the later claims of another extensor mechanism disruption and revision and/or two stage revision thereafter.No additional materials were presented for review.Event confirmation: two arthroscopy procedures can be confirmed.A revision for a loose femur can be confirmed.A second revision for a loose femur and fracture of a femoral stem can be confirmed.The reason for the stem fracture cannot be confirmed.The two staged revision to competitor components cannot be confirmed.Root cause: the root cause for arthroscopy #1 was anterior knee pain, it band tendonitis and patellar tilt.The cause of the first femoral revision was loosening but a root cause cannot be attributed.The root cause of the second arthroscopy was pain and swelling and presence of metallic debris in aspirated fluid.The root cause of the metal stained fluid was motion at a fractured femoral stem.The root cause of the second revision was femoral loosening and a fractured femoral stem.The root cause of the stem fracture cannot be determined.The root cause of the last procedure, wash out and polyethylene revision, was infection.The root cause of the two stage revision which was not documented would presumed to be infection.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: a review of the provided medical information by a clinical consultant indicated: there were revision surgeries for femoral loosening and patellar mal-tracking but the work-up and etiology for these were not clear.The last revision was result of a failure of the femoral component-stem construct at the threaded junction.The likely cause of the failure was micromotion at the threaded junction led to wear and material fatigue.To determine the mode of failure additional materials would be required as well as evaluation of the explant.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device is needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
In performing follow-up, rep provided a revision operative report for (b)(6) 2013 for the patient's right knee.Preoperative diagnosis: loose right revision knee arthroplasty with maltracking of patella.Op report notes: "in less than 1 year she developed increasing pain and discomfort.Studies are suspicious for possible loosening of femoral component.There was only some distal cement around the femoral component.The stem appeared to be reactive line around it suggesting loosening of femoral component.The patella was also tilting and maltracking¿femoral component came out very easily with no fixation at all.The cement was bonded to the implant with augments.The stem came out and there was a fibrous membrane around the stem completely.There was no cement on the stem at all consistent with femoral component loosening.We had to remove the patellar component to position it more centrally.".
|
| |
|
Search Alerts/Recalls
|