Model Number MMT-105NNPKNA
Device Problem Computer Software Problem (1112)
Patient Problem Hypoglycemia (1912)
Event Date 02/05/2020
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.This mdr is related to the retrospective review of complaints from companion medical inc, following medtronic¿s acquisition of the company in (b)(6) 2020.
 
Event Description
Customer reported via phone call that there was inaccuracy between the dose delivered and the dose recorded in the app.Customer stated the discrepancy was greater than 1.0 unit.Customer went low and took more insulin to make up for the lower dose recorded in the app.No harm requiring medical intervention was reported.The device was not returned for analysis.
 
Manufacturer Narrative
Pen paired to samsung s10 successfully.A test dose of 6.5u was dialed and dosed 10 times.The logbook displayed 6.5 each time and without delay.Patient firebase data dose duration time on (b)(6) 2020 suggests the patient dosed too slowly resulting in a larger dose split into two smaller prime doses.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105NNPKNA NOVO NORDISK PINK
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
MDR Report Key11901578
MDR Text Key255961933
Report Number3012822846-2021-00296
Device Sequence Number1
Product Code FMF
UDI-Device Identifier000010862088000368
UDI-Public(01)000010862088000368(10)A92HC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 10/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-105NNPKNA
Device Catalogue NumberMMT-105NNPKNA
Device Lot NumberA92HC
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2020
Date Manufacturer Received10/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-