• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL LP BONE CEMENT; COBALT HV BONE CEMENT 40GM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL LP BONE CEMENT; COBALT HV BONE CEMENT 40GM Back to Search Results
Model Number 600-15-000
Health Effect - Clinical Codes Hematoma (1884); Pain (1994); Swelling/ Edema (4577)
Date of Event 12/15/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Revision surgery - patient had a total knee arthroplasty with cobalt bone cement.Patient experienced swelling, pain and a hematoma.Patient subsequently sought treatment from another provider and had revision surgery.The femur was reported to be loose.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as swelling, pain, hematoma and loose femur.The previous surgery and the surgery detailed in this event occurred 2 years apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There was no non-conforming material report (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to swelling, pain, hematoma and loose femur.There were no findings during this evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE CEMENT
Common Device Name
COBALT HV BONE CEMENT 40GM
Manufacturer (Section D)
ENCORE MEDICAL LP
9800 metric blvd
austin TX 78758 5445
MDR Report Key11901604
Report Number1644408-2021-00414
Device Sequence Number4175914
Product Code LOD
UDI-Device Identifier00190446138006
UDI-Public(01)00190446138006
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K051496
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/16/2020
Device Model Number600-15-000
Device Catalogue Number600-15-000
Device Lot Number820D2D0011
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/03/2021
Supplement Date Received by Manufacturer05/28/2021
Initial Report FDA Received Date05/28/2021
Supplement Report FDA Received Date06/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
-
-