ORTHALIGN, INC. ORTHALIGN PLUS SYSTEM; REFERENCE SENSOR 5 Back to Search Results
Model Number 403087
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
At this time the product has not been returned for investigation.Once the product is returned orthalign will perform an investigation into the alleged accuracy issue.Orthalign is filing this mdr with an abundance of caution with the understanding of the potential harm that the patient could be subject to if the device produces inaccurate measurements.
 
Event or Problem Description
It was reported cup angle values were opposite/different than what was expected.Went through checklist of potential reasons, and all was done correctly.
 
Additional Manufacturer Narrative
The root cause according to the nav.Unit log file is most likely due to the rs5 sensor being unstable in communication check and the data received being abnormal.The navigation unit passed all related functional tests and functioned as designed when used during in-house functional testing with an in-house rs5.A review of the device history record (dhr) was conducted.The device passed all manufacturing specifications prior to release.Orthalign, inc.Will continue to monitor this issue and take action when/if alert limits are exceeded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORTHALIGN PLUS SYSTEM
Common Device Name
REFERENCE SENSOR 5
Manufacturer (Section D)
ORTHALIGN, INC.
120 columbia
suite 500
aliso viejo CA 92656
MDR Report Key11904394
Report Number3007521480-2021-00006
Device Sequence Number3551167
Product Code OLO
UDI-Device Identifier00810832030072
UDI-Public00810832030072
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K172462
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number403087
Device Catalogue Number403087-06
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2021
Supplement Date Received by Manufacturer04/28/2021
Initial Report FDA Received Date05/28/2021
Supplement Report FDA Received Date08/31/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-