SMITH & NEPHEW, INC. GII PS HI FLEX ISRT SZ 3-4 9; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Model Number |
71421508 |
| Medical Device Problem Code |
Unstable (1667)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during surgery, when installing a gii ps hi flex isrt sz 3-4 9, it was unstable.Device inside the patient.The procedure was completed using a s+n back-up device.Surgery was less than or equal to 30mins delayed.Patient was not harmed.
|
| |
|
Additional Manufacturer Narrative
|
|
A2: patient age available.G3, h2,h3 and h6: the device, used in treatment, was returned for evaluation.A visual inspection of the returned device confirms damage on the lock detail probably from the attempted insertion.The visual also confirms some scratches on the surface of the insert.A dimensional inspection was attempted but the device was too damaged from the attempted insertion to obtain accurate measurements.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.D8, d9: device returned g2: report source update.
|
| |
|
Search Alerts/Recalls
|
|
|