Model Number 71332754
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 05/17/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after a thr was performed on (b)(6) 2017, a revision surgery was performed on (b)(6) 2021 tto removed the 0 deg xlpe acet lnr 36mm x 54mm ((b)(4)) and oxinium fem hd 12/14 36 mm +0 ((b)(4)) due to dislocation, and replaced them with an or3o insert and liner with a 28mm +8 oxinium head on.No patient harm beyond the described event.
 
Additional Manufacturer Narrative
H6: the device, used in treatment, was not returned for evaluation.The pictures provided were reviewed and could not confirm the stated failure mode.The clinical/medical investigation concluded that, the two undated, unlabeled x-rays were provided but don¿t contribute to the root cause.Based on the limited information provided, the root cause of the reported dislocation cannot be determined.The impact to the patient beyond that which has already been reported, cannot be determined.Should any additional medical information be provided, this complaint will be re-assessed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, patient anatomy or abnormal loading of limb.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 0 DEG XLPE ACET LNR 36MM X 54MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11908918
Report Number1020279-2021-04834
Device Sequence Number14575554
Product Code JDI
UDI-Device Identifier03596010597991
UDI-Public03596010597991
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K113848
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71332754
Device Catalogue Number71332754
Device Lot Number17JM15428
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/17/2021
Supplement Date Received by Manufacturer07/30/2021
Initial Report FDA Received Date05/31/2021
Supplement Report FDA Received Date08/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-