• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. 11.5 X 40 130 D TAN LT LIME; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. 11.5 X 40 130 D TAN LT LIME; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71647340
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 05/06/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a nail surgery had been performed on an unknown date, the patient experienced pain at the screw sites (part and lot number will not be available).The issue was addressed via removal of 11.5 x 40 130 d tan lt lime in a revision surgery on (b)(6) 2021.The removal was quick and without incident.It is unknown if there was a replacement of instrument.
 
Additional Manufacturer Narrative
H6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, no relevant supporting clinical information has been provided to assist with a clinical investigation.The patient's current condition is unknown and the patient impact beyond the reported events could not be determined based on the limited information provided.Therefore, based on insufficient information, a thorough clinical assessment cannot be performed at this time.Should any additional clinical information be provided this complaint will be re-evaluated.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, joint tightness, material in use, patient reaction or loss of sterility.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11.5 X 40 130 D TAN LT LIME
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11912857
Report Number1020279-2021-04877
Device Sequence Number3633229
Product Code HSB
UDI-Device Identifier03596010500946
UDI-Public03596010500946
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K040462
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/04/2021
Device Model Number71647340
Device Catalogue Number71647340
Device Lot Number11AM03005
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 05/06/2021
Supplement Date Received by Manufacturer08/09/2021
Initial Report FDA Received Date06/01/2021
Supplement Report FDA Received Date08/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
-
-