Model Number 9735669 |
Device Problems
Device Sensing Problem (2917); Mechanics Altered (2984)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Onsite functional and visual examination was performed by a manufacturer representative.The emitter was replaced and the issue was resolved.The system passed a system checkout and was returned to an operational condition.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation device being used for a functional endoscopic sinus surgery (fess).It was reported that the site's break out box (bob) port 3 and 5 have a loose connection and are intermittent with nipt tracking.The nipt consistently tracks if the plug in portion is held in place, once they let go the navigation is in and out with those two ports only.Different nipt were used with same results.Surgery continued with nipt in other ports.There was no reported delay and no reported impact on patient outcome.
|
|