| Model Number |
PARADYM RF CRT-D 9750 |
| Medical Device Problem Code |
Reset Problem (3019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reportedly, upon device interrogation on (b)(6) 2021, a warning stating that a reset occurred on (b)(6) 2021 and re-initialization was performed was displayed.The last battery reforming was performed on (b)(6) 2021.On (b)(6) 2021, the battery voltage was at 2.88v.A charge time of 15s was measured.The device was successfully reprogrammed and the statistics were reset.Proper device operation was observed.The patient had undergone several surgeries since the beginning of the year.
|
| |
|
Event or Problem Description
|
|
Reportedly, upon device interrogation on (b)(6) 2021, a warning stating that a reset occurred on (b)(6) 2021 and re-initialization was performed was displayed.The last battery reforming was performed on (b)(6) 2021.On (b)(6) 2021, the battery voltage was at 2.88v.A charge time of 15s was measured.The device was successfully reprogrammed and the statistics were reset.Proper device operation was observed.The patient had undergone several surgeries since the beginning of the year.
|
| |
|
Additional Manufacturer Narrative
|
|
Please refer to the attached analysis report.
|
| |
|