| Model Number |
7213 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the returned product consisted of an nc emerge balloon catheter.The device was microscopically and visually examined.The hypotube of the device has numerous kinks.There was a separation 56.7cm from the strain relief.There was both contrast and blood in the inflation lumen and balloon.There was contrast present in the guidewire lumen.The balloon was tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon detachment occurred.The 95% stenosed target lesion was located in the non-tortuous and moderately calcified proximal left anterior descending artery.A 4.00mm x 15mm nc emerge balloon catheter was selected for use.During insertion of the balloon catheter through the guider, the balloon never made it past the curve of arch.The proximal half was removed as normal through the touhy and visualized under fluoro prior to removal of guider and guidewire.The guider was flushed which did not eject portion of balloon catheter.The distal end remained in the guider and guide/guidewire.A guidewire was inserted through the guider which pushed the remainder of the balloon catheter out of guider.The remaining portion were removed as one unit.The two portions of balloon catheter were inspected and appeared intact; the shear was in mid catheter.The device was completely removed and the procedure was completed with a new 4x15 nc emerge balloon catheter.No patient complications were reported.
|
| |
|
Search Alerts/Recalls
|