|
During investigation, two cut-like cracks were detected at the transition to the pump connector.Crack 1 is approx.8 mm in the circumferential direction while crack 2 runs in the longitudinal direction and separates the driving tube approx.1.5 cm from the end.Both cracks were identified as cuts.It is very likely that the longitudinal cut (crack 2) was intentionally made in order to be able to separate the driving tube from the pump connection after identifying the first crack.The circumferential cut (crack 1) was probably caused by a sharp object.In the current instructions for use (ifu), the user is advised to ensure that the cannulae and driving tubes are not subjected to any tension, torsion, pressure, or traction.Also, the user is advised against touching the cannulae and driving tubes with sharp-edged objects, that can do damage to the material.And further on, in the manual the user of excor is instructed to inspect the blood pumps, cannulae and driving tubes regularly.
|