MEDTRONIC NEUROSURGERY STRATA NSC LUMBOPERITONEAL SHUNT KIT; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 44420
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that when the guide wire was used to reinforce the toughness of the lumbosacral end, breakage occurred.
 
Event or Problem Description
Additional information received reported that there were no environmental/external/patient factors that may have led or contributed to the issue.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
The returned guidewire appeared to have a sharp bending at the center of the wire.It is unknown how this damage may have occurred.Dhr review of lot # did not indicate any defects related to the failure seen in the field.Ifu cautions the user, "if using a guidewire, remove the tuohy needle and guidewire from the catheter simultaneously.Hold the catheter securely while the guidewire is being removed, being careful not to pinch the catheter tightly.Remove the guidewire slowly and evenly.Removing the guidewire too quickly can slice the catheter from inside out." medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATA NSC LUMBOPERITONEAL SHUNT KIT
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
5290 california ave
irvine CA 92617
MDR Report Key11913532
Report Number2021898-2021-00124
Device Sequence Number17107791
Product Code JXG
UDI-Device Identifier00643169975866
UDI-Public00643169975866
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K091312
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model Number44420
Device Catalogue Number44420
Device Lot NumberE74363
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/29/2021
Supplement Date Received by Manufacturer06/02/2021
08/12/2021
Initial Report FDA Received Date06/01/2021
Supplement Report FDA Received Date06/08/2021
08/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-