Model Number ROB10024
Device Problem Application Program Problem: Parameter Calculation Error (1449)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/19/2021
Event Type  malfunction  
Event Description
It was reported that during a cori assisted tka surgery, the real intelligence cori received a failure to generate bone model error.They cleared a few tibial points, hit continue, and the bone model generated.Case proceeded as planned.The procedure was completed, with a minimal delay (fewer than 30 minutes) using the same device.No patient injuries and no other complications were reported.
 
Manufacturer Narrative
Section h3, h6: the real intelligence cori, pn: rob10024, sn: (b)(6) used for treatment was not returned to the designated complaint unit for evaluation, visual and functional inspections could not be performed.Screenshots of the case were provided and confirmed the warning message upon entering the tibia bone removal stage: ¿bone model generation error¿.This is a known software issue that has been corrected and released in cori-v1.4.3 software.Refer to the real intelligence cori for knee arthroplasty user manual for data point collection, including tibia free collection.If a collection error occurs, a message displays, reporting the specific error to the user, and providing instruction for solution.Most indications require clearing the message from the screen, by pressing ok, and performing the steps again to re-collect.This situation is captured in the risk assessment released at the time of the complaint.The failure mode and associated risk have been anticipated within the risk file and that the documented risk level is still adequate.A review of manufacturing records indicate the software met all specifications upon release into distribution.A complaint history review found similar reports.Escalation actions applicable to the scope of the reported complaint have been identified, and it was determined that no further action is required at this time.This issue will be continuously monitored through complaint investigation and post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE CORI
Type of Device
REAL INTELLIGENCE CORI
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11913803
MDR Text Key253535465
Report Number3010266064-2021-00430
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757420
UDI-Public00885556757420
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10024
Device Catalogue NumberROB10024
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-