Model Number HYPERTHERMIA PUMP
Medical Device Problem Code Intermittent Loss of Power (4016)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/03/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The hyperthermia pump involved in the incident has not been returned to belmont for investigation, therefore a root cause of the reported power loss cannot be established.The manufacturing and service records for this serial number were reviewed and no anomalies were identified.The unit has not been returned to belmont for service or preventive maintenance since it was shipped to the customer in 2017.Without the ability to confirm the complaint, it is not possible to establish a root cause of the reported power loss.It was reported that another hyperthermia pump was used to finish the case; there was no injury to the patient.Belmont will continue to monitor and trend similar reports of this nature.Should the unit be returned for investigation, or additional information becomes available, a supplemental report will be provided.
 
Event or Problem Description
The user facility reported that the hyperthermia pump was frequently shutting down a case.Another hyperthermia pump was brought in to finish the case.There was no harm to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THE BELMONT HYPERTHERMIA PUMP
Common Device Name
THERMAL INFUSION FLUID WARMER
Manufacturer (Section D)
BELMONT MEDICAL TECHNOLOGIES
780 boston road
billerica MA 01821
MDR Report Key11917100
Report Number1219702-2021-00073
Device Sequence Number3710233
Product Code LGZ
UDI-Device Identifier10896128002456
UDI-Public(01)10896128002456
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K152208
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 06/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHYPERTHERMIA PUMP
Device Catalogue Number902-00001
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/04/2021
Initial Report FDA Received Date06/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-