MAUDE Adverse Event Report: INSULET CORPORATION OMNIPOD INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSULET CORPORATION OMNIPOD INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN Back to Search Results
Model Number 18900-5C
Device Problem Failure to Read Input Signal (1581)
Patient Problem Hyperglycemia (1905)
Event Date 05/26/2021
Event Type  Injury  
Manufacturer Narrative
The device has not been returned/received to date.If the device is received, a supplemental report will be submitted with the investigation results.We are unable to determine if any product condition could have contributed to the reported er visit and hyperglycemia.Lot release records were reviewed and the product lot met all acceptance criteria.Specifically, a pod is paired to a pdm and put through simulated use testing including communicating with the pdm, deployment, delivering fluid, occlusion detection, and freedom from hazard alarms.
 
Event Description
It was reported that the patient visited the emergency room (er) due to high blood glucose (bg) levels of 18.3 mmol/l (329.4 mg/dl) with ketones present.The patient stated that the omnipod personal diabetes manager (pdm) is not working properly when inputting the blood readings to correct and deliver bolus.At the hospital, the patient was given insulin through a manual injection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
jordan biswurm
100 nagog park
acton, MA 01720
9786007000
MDR Report Key11917888
MDR Text Key253802969
Report Number3004464228-2021-08554
Device Sequence Number1
Product Code LZG
UDI-Device Identifier10385081120067
UDI-Public(01)10385081120067(11)180717(10)L60489
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192659
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 05/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number18900-5C
Device Catalogue NumberENT450
Device Lot NumberL60489
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/26/2021
Initial Date FDA Received06/01/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/17/2018
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age35 YR
-
-