COOK ENDOSCOPY ACUSNARE POLYPECTOMY SNARE SOFT; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
|
Back to Search Results |
|
| Model Number |
G22703 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter section: occupational - manager.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use states: "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use." instructions for use states: "fully retract and extend snare to confirm smooth operation of device." instructions for use states: "inspect active cord.Cord must be free of kinks, bends, breaks, and exposed wires to allow for accurate transfer of current.If an abnormality is noted, do not use active cord." prior to distribution, all acusnare polypectomy snares are subjected to a visual inspection and functional testing to ensure device integrity.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered remote.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
During a polypectomy, the physician used a cook acusnare polypectomy snare soft.The catheter was crimping [kinking] near the handle when they go around the polyp, so it does not cut properly.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Search Alerts/Recalls
|
|
|