SIEMENS HEALTHCARE GMBH SENSIS HEMO LOW; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 6634633
Device Problem Intermittent Communication Failure (4038)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/27/2021
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that a malfunction occurred while operating the sensis hemo low system.During an interventional procedure, the user reported a communication problem between the artis and sensis systems.The procedure was continued and completed on an alternate system.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a software error.The detailed investigation showed that the problem was due to a software error.This caused temporary problems with patient registration on site, but this was resolved by restarting the system.The above is an issue for which a corrective action already exists.The corresponding software update is sw update vc12m patch 2 (ax022/19/p) and is already available.According to our records, the required patch has been applied on site.Since then, the error has not been reported again.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSIS HEMO LOW
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
siemensstr. 1
forchheim, germany 91301
GM  91301
MDR Report Key11921030
MDR Text Key255889063
Report Number3004977335-2021-82758
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
PMA/PMN Number
K150493
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Repair
Type of Report Initial,Followup
Report Date 06/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6634633
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-