MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE STRYDE SYSTEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE STRYDE SYSTEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number PS10.0-65D250
Medical Device Problem Code Material Erosion (1214)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information has been provided, a supplemental report will be submitted.
 
Event or Problem Description
Information was received that the patient experienced pain at the junction of the nail.Upon removal, it was noted that there was rust at the junction and a huge callus formation.
 
Event or Problem Description
No additional information was provided.
 
Additional Manufacturer Narrative
The nail was returned for visual and functional testing.Visual inspection revealed that the device had corrosion on the junction of the distraction rod.Functional testing was not applicable due to the reported failure mode.Device history review: a device history review (dhr) was performed on lot number: 9051314 and there were no deviations in the manufacturing process and the finished product met all acceptance criteria prior to shipment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECICE STRYDE SYSTEM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
MDR Report Key11922080
Report Number3006179046-2021-00288
Device Sequence Number19915623
Product Code HSB
UDI-Device Identifier00887517951175
UDI-Public887517951175
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K180503
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPS10.0-65D250
Device Lot Number9051314
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/20/2021
Supplement Date Received by Manufacturer08/13/2021
Initial Report FDA Received Date06/02/2021
Supplement Report FDA Received Date08/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age14 YR
Patient Weight60
-
-