• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN 2.5MM HEX SCRWDRVR; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN 2.5MM HEX SCRWDRVR; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71434455
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/21/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during field inspection, it was found that two (2) lgn 2.5mm hex scrwdrvr are stripped.No case involved; therefore, there was no patient involvement.
 
Additional Manufacturer Narrative
H3, h6: the associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was stripped along the tip, rendering the device inoperative.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN 2.5MM HEX SCRWDRVR
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11922446
Report Number1020279-2021-04964
Device Sequence Number3268479
Product Code HRY
UDI-Device Identifier03596010550200
UDI-Public03596010550200
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71434455
Device Catalogue Number71434455
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2021
Initial Date Received by Manufacturer 05/21/2021
Supplement Date Received by Manufacturer07/29/2021
Initial Report FDA Received Date06/02/2021
Supplement Report FDA Received Date07/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-