Model Number 0684-00-0576-01 |
Device Problems
Off-Label Use (1494); Optical Problem (3001); Pressure Problem (3012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/11/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
|
|
Event Description
|
It was reported that during intra-aortic balloon (iab) therapy, a fiber optic sensor failure occurred.The insertion was reported to be axillary, which is not the method described in the device instructions for use.The customer attempted to re-connect the iab without success.It was noted that there was a waveform present, but no blood pressure (bp) numbers were displayed.The customer then transduced the central lumen and zeroed the transducer.There was no patient harm or adverse event reported.
|
|
Manufacturer Narrative
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.The device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.The review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.The overall complaint trend data for the period may-19 through apr-21 was reviewed.There were no triggers identified for the review period.Communication/interviews were performed to obtain all possible information.Reference complaint #(b)(4).
|
|
Search Alerts/Recalls
|