• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG CASPAR RONGEUR UP-BITE 3MM155MM; BONE PUNCHES, RONGEURS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG CASPAR RONGEUR UP-BITE 3MM155MM; BONE PUNCHES, RONGEURS Back to Search Results
Model Number FF843R
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Insufficient Information (4580)
Date of Event 03/07/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event or Problem Description
It was reported that there was an issue with ff843r - caspar rongeur up-bite 3mm155mm.According to the complaint description, a rongeur with reference ff843r is broken during a spine surgery on (b)(6) 2017.A piece of the rongeur remained in the body of the patient.A revision surgery was necessary.Additional information received confirmed that the specific piece of the rongeur has been removed (d.D.End (b)(6)/beginning of (b)(6)).The adverse event is filed under (b)(4).
 
Additional Manufacturer Narrative
Updated e1: initial reporter.Updated h6: codes.Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: due to the fact that no lot number was provided, a review of the device history records for the complained device is not possible.The review of risk assessment revealed that the overall risk level (severity 4(5) x probability of occurrence 2(5)) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CASPAR RONGEUR UP-BITE 3MM155MM
Common Device Name
BONE PUNCHES, RONGEURS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key11923458
Report Number9610612-2021-00417
Device Sequence Number804896
Product Code HTX
UDI-Device Identifier04038653044719
UDI-Public4038653044719
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFF843R
Device Catalogue NumberFF843R
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2021
Supplement Date Received by Manufacturer09/06/2021
Initial Report FDA Received Date06/02/2021
Supplement Report FDA Received Date09/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-