| Model Number |
3243 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.Initial reporter city: (b)(6).
|
| |
|
Event or Problem Description
|
|
It is reported that a rotapro and rotawire entrapment occurred.A 1.75mm rotapro and a 300cm rotawire were selected for use in a percutaneous coronary intervention (pci) procedure.The 90% stenosed target lesion was moderately tortuous and severely calcified right coronary artery (rca).The device was prepared outside the body and platformed at 180,000rpm, then advanced over the wire.During procedure, the burr became stuck on the wire inside the patient.The devices were removed together as one unit, and severe resistance was felt during removal.After it was removed, the physician noticed the rotawire was kinked.The procedure was completed successfully with this device and there were no patient complications reported.
|
| |
|
Event or Problem Description
|
|
It is reported that a rotapro and rotawire entrapment occurred.A 1.75mm rotapro and a 300cm rotawire were selected for use in a percutaneous coronary intervention (pci) procedure.The 90% stenosed target lesion was moderately tortuous and severely calcified right coronary artery (rca).The device was prepared outside the body and platformed at 180,000rpm, then advanced over the wire.During procedure, the burr became stuck on the wire inside the patient.The devices were removed together as one unit, and severe resistance was felt during removal.After it was removed, the physician noticed the rotawire was kinked.The procedure was completed successfully with this device and there were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A2: age at time of event: 18 years or older.E1: initial reporter city: (b)(6).Device evaluated by mfr: the returned product consisted of the rotapro atherectomy system and the burr catheter was attached to the advancer unit.The advancer, handshake connections, sheath, coil, and burr were visually and microscopically examined.Inspection of the device presented no damage or irregularities.The wire used in the procedure was returned, so functional testing was completed with the returned rotawire.The wire was able to be inserted into the burr and advanced through the advancer with no resistance or issues.
|
| |
|