| Model Number |
394919 |
| Medical Device Problem Code |
Premature Elective Replacement Indicator (1483)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Device explanted due to eri.No adverse patient events reported.Should additional information become available, it will be added to this file.
|
| |
|
Additional Manufacturer Narrative
|
|
We received your event description for the above mentioned device and would like to thank you for supporting our post-market surveillance.As of today, the medical device is not available for analysis, therefore the device itself could not be investigated.The information you provided has been entered into our quality system as a complaint.These types of complaints are used to evaluate systems and device performance throughout our organization and help to maintain and improve the performance of our devices.Should additional relevant information or the device itself become available, the investigation will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for analysis.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated and all production steps were performed accordingly.Particularly the final acceptance test proved the device functions to be as specified.Upon receipt, the pacemaker was interrogated and the memory content was analyzed.The battery status showed 40 percent.The archive data and the programmed parameters were inspected.An elevated current consumption was documented in the archive data due to high left ventricular pacing outputs of 5.8 v and 1.5 ms as well as high right ventricular pacing outputs of 3.0 v and 0.4 ms.This represents a high current program, which results in a faster discharge of the battery.The amount of charge taken from the battery was verified and the battery condition was found to be as expected.The pacemaker was subjected to an electrical analysis.The ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.In conclusion, the device proved to be fully functional.The analysis of the memory content showed a normal device behavior while the device was implanted and in service.There was no indication of a device malfunction.
|
| |
|
Search Alerts/Recalls
|