| Model Number |
7015 |
| Medical Device Problem Codes |
Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter address: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The 90% stenosed target lesion was located in the none calcified and non tortuous mid of left anterior descending artery.Following stent implantation, a 2.5mm x 15mm quantum maverick balloon catheter was advanced for post-dilatation.However, during expansion the physician found the front end of the catheter was broken, 40 cm from the hub but not completely fractured.The device was completely removed in a normal manner and the procedure was successfully completed with a different device.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The 90% stenosed target lesion was located in the none calcified and non tortuous mid of left anterior descending artery.Following stent implantation, a 2.5mm x 15mm quantum maverick balloon catheter was advanced for post-dilatation.However, during expansion the physician found the front end of the catheter was broken, 40 cm from the hub but not completely fractured.The device was completely removed in a normal manner and the procedure was successfully completed with a different device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
E1.Initial reporter address: (b)(6).Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter.The device was microscopically and visually examined.There was a hypotube separation 92.8cm from the strain relief.At 90cm from the strain relief the device was kinked.There was exit notch damage to the device.There was contrast and blood in the inflation lumen and balloon folds.The balloon was tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported break as there was separation to the device.
|
| |
|