MAUDE Adverse Event Report: ADVANCED STERILIZATION PRODUCTS STERRAD 100NX STERILIZER ALLCLEAR 1-DR; STERRAD® EQUIPMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED STERILIZATION PRODUCTS STERRAD 100NX STERILIZER ALLCLEAR 1-DR; STERRAD® EQUIPMENT Back to Search Results
Model Number 10104-005
Medical Device Problem Code Environmental Particulates (2930)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A field service engineer was dispatched to the customer site.The catalytic converter, oil mist filter and vacuum pump oil were replaced to resolve the smoke/haze issue.Unit meets specifications and was returned to service.Asp complaint ref #: (b)(4).
 
Event or Problem Description
A customer reported an event of a "smoke" or haze emitting from the sterrad® 100nx sterilizer.There was no report of any injuries or human reactions.An asp field service engineer was dispatched to assess the unit onsite.This event is being reported as a malfunction report subsequent to a serious injury.
 
Additional Manufacturer Narrative
H3: asp investigation summary: the investigation included a review of the device history record (dhr), trending analysis of the smoke/haze issue, and system risk analysis (sra).¿the device history record (dhr) was reviewed and no issues relating the failure mode were noted.The involved unit met manufacturer specifications at the time of release.¿trending analysis of the smoke/haze issue for the sterrad® 100nx unit was reviewed for the prior six months from open date and no significant trend was observed.¿review of risk documentation shows the risk for exposure to toxic or corrosive material to be "low." the catalytic converter, oil mist filter, and vacuum pump oil were not returned for further evaluation.The assignable cause of the smoke/haze is likely due to the catalytic converter, oil mist filter, and vacuum pump oil.The asp field service engineer replaced the parts and confirmed the sterrad® 100nx was restored to proper function after service.The issue was resolved at the customer facility.Asp will continue to track and trend this issue.Asp complaint ref #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERRAD 100NX STERILIZER ALLCLEAR 1-DR
Common Device Name
STERRAD® EQUIPMENT
Manufacturer (Section D)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
MDR Report Key11925908
Report Number2084725-2021-00187
Device Sequence Number3197438
Product Code MLR
UDI-Device Identifier10705037048670
UDI-Public10705037048670
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K071385
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 08/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10104-005
Device Catalogue Number10104-005
Device Lot NumberN/A
Was the Report Sent to FDA? Yes
Initial Date Received by Manufacturer 05/20/2021
Supplement Date Received by Manufacturer08/12/2021
Initial Report FDA Received Date06/02/2021
Supplement Report FDA Received Date08/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-