| Model Number |
2346-01-000 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Broken white t-handle, dull calcar planar, stripped liner trial, cracked cup inserter, stripped screwdriver no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned.The investigation could not verify or identify any product contribution to the reported event with the information provided.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history review: a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
|
| |
|
Search Alerts/Recalls
|