Model Number 9697
Medical Device Problem Code Missing Test Results (3267)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/07/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient identifier: so personal information was not entered.No information was captured as the customer's weight was not provided.
 
Event or Problem Description
The customer reported that her blood glucose readings were not being stored in the memory of the contour next meter.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.Replacement meter kit and test strips were sent to the customer.
 
Additional Manufacturer Narrative
The customer returned the suspected contour next meter and contour next test strips for evaluation.However, the returned test strips were expired.Therefore, the returned meter was tested with the in-house contour next test strips from a different lot # than one associated with the event using a blood sample, which gave a satisfactory performance.The blood glucose readings obtained during in-house testing were properly stored in the meter's memory.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR NEXT
Common Device Name
BLOOD GLUCOSE METER
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
100 summit lake drive
valhalla NY 10595
Manufacturer (Section G)
PHC CORPORATION (REGISTRATION #: 3004167884)
110 oaza-inoshiri-aza
-nishiueno, wakimachi
mima tokusima 779-3 603
JA   779-3603
Manufacturer Contact
shweta gulati
100 summit lake drive
valhalla, NY 10595
9142361830
MDR Report Key11929468
Report Number1810909-2021-00235
Device Sequence Number17242207
Product Code NBW
UDI-Device Identifier00301939697010
UDI-Public00301939697010
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K121190
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 05/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number9697
Device Catalogue Number9697
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/07/2021
Supplement Date Received by Manufacturer05/12/2022
Initial Report FDA Received Date06/03/2021
Supplement Report FDA Received Date05/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/24/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age83 YR
Patient SexFemale
-
-