| Model Number |
041210002A |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/20/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Sample expected to return for evaluation.Follow-up report will be submitted once the sample has been received and reviewed.
|
| |
|
Event or Problem Description
|
|
Argon 4-way stopcock was cracked, causing fluid to leak from iv hookup.No patient harm.Used:tubing /iv sets.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of one stopcock assembly found severe damage/cracks to the luers which would result in leakage, confirming the complaint.Root cause investigation found no discrepancies with the molding process or the annealing process which could be factors in the failure observed.Additional testing was conducted on in-house parts, pre and post anneal process, and the damage/cracks could not be duplicated.The cause of the damage/cracks may possibly be related to the type of substance used to wipe the product or that was injected through the product.The ifu warns, ¿do not use patient monitoring products with parenteral nutritional solutions containing lipids or nitroglycerin.These substances are known to degrade some medical plastics resulting in cracks and/or leaks.Other drugs or solutions may have similar effects¿.The user facility has not responded to requests for additional information to confirm if incompatible substances were used on the product.
|
| |
|
Event or Problem Description
|
|
Argon 4-way stopcock was cracked, causing fluid to leak from iv hookup.No patient harm.Used:tubing /iv sets.
|
| |
|