| Model Number |
S-50-100-120-P6 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/20/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents and/or complaints reported from this lot.The investigation was unable to determine a cause for the reported entrapment and tip separation resulting in unexpected medical intervention.It may be possible that the thumbslide was not retracted to the start position and the system lock was not locked prior to removing the delivery system causing the tip to catch on the stent, anatomy or introducer sheath during removal; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a superficial femoral artery.Post-stent implantation, while removing the delivery system of a 5x100mm supera resistance was met with the guide wire when the nose cone separated.The supera delivery system was able to be removed and then a snare was used to retrieve the separated tip.There were no adverse patient sequela and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|