SMITH & NEPHEW, INC. RDCE FRCPS W/ PTS BRD; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
Model Number 71173377 |
Device Problem
Defective Device (2588)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device, intended for use in treatment, was returned for evaluation. a visual inspection was performed and confirmed the forceps are bent at the tip. a review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the complaint history for the listed part revealed prior complaints for the listed failure mode with the same batch number.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Event Description
|
It was reported that, the clamp of a rdce frcps w/ pts brd was observed to be damaged (b)(4) and that the tip of a t8 linear driver shaft w/ ao qc is twisted (b)(4).It is unknown whether the event happened during surgery and if there was a patient involvement.During investigation approved on (b)(6) 2021 was confirmed that forceps are bent at the tip.
|
|
Search Alerts/Recalls
|
|
|