| Model Number |
1217-01-056 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Synovitis (2094)
|
| Date of Event |
10/03/2011
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Unf and medical records received.After review of medical records patient was revised to addressed left failed total hip arthroplasty with squeaking metal on metal articulation.Once we entered the joint there was release of darkly fluid.Thick gray reactive synovitis with multiple areas of black dense material scattered through the synovium.Metallosis was tracked along the acetabular and neck of femoral component bone interface.It was obvious that the acetabular component was too vertical.Doi: (b)(6) 2005 dor: ((b)(6) 2011 left hip.Please see (b)(4) for right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the material, manufacturing, inspection or sterile processing that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|