Model Number AP2-000002
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Seaspine was made aware of this event on 4 may 2021.The drivers were received for analysis, and visual inspection confirmed damage to the tips.Functional torque testing of similar screws and screwdrivers made from the same material and material condition was conducted.During this test, no driver failed, although wear was noted.While the recorded user applied loads remained under the screwdriver ultimate torque, driver wear was noted when the driver was cyclically loaded to high torque values.Therefore, root cause is not determined but may be due to cyclic loading followed by a high torque application.Review of labeling: possible adverse events: bending, disassembly or fracture of implant and components.
 
Event or Problem Description
The patient underwent spinal surgery consisting of seaspine's admiral acp system on (b)(6) 2021.During the surgery, the ap2-000002 screwdriver, threaded tip broke, and a fragment remained in the screw head.A second ap2-000002 screwdriver, threaded was provided, and this instrument also broke in a similar manner, also leaving a fragment in a screw head (reference #3012120772-2021-00048).The surgeon determined the fragments could remain in the screw heads, as the fragments were below the head surface, and the locking cover placed over the screw further mitigated the risk of migration.The surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADMIRAL ACP SYSTEM
Common Device Name
SCREWDRIVER, THREADED
Manufacturer (Section D)
SEASPINE INC.
5770 armada drive
carlsbad CA 92008
MDR Report Key11935341
Report Number3012120772-2021-00047
Device Sequence Number3182018
Product Code KWQ
UDI-Device Identifier10889981239190
UDI-Public10889981239190
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K202064
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAP2-000002
Device Catalogue NumberAP2-000002
Device Lot NumberCR42755B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/06/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2021
Initial Report FDA Received Date06/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-