Model Number 72290143
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/11/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopy, the wire of the double fix knotless threader was broken, the device could not be used.The device broke inside the patient and all the pieces were removed.It is unknown how were they were removed.The procedure was completed with a s+n back up device.No delay and no other complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection of the instrument shows no manufacturing abnormalities.The snare wire and cross bar were returned with device damaged.A basic functional evaluation shows the deployment knob can be rotated and stop clockwise.The suture loading portion of the device was returned damaged and cannot be functionally tested.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.The complaint was confirmed.Factors that could have contributed to the reported event include: (1) excessive force.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOUBLE FIX KNOTLESS PK
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key11936067
Report Number3006524618-2021-00692
Device Sequence Number1297096
Product Code MBI
UDI-Device Identifier00885556722497
UDI-Public00885556722497
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153669
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/04/2022
Device Model Number72290143
Device Catalogue Number72290143
Device Lot Number2045342
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2021
Initial Date Received by Manufacturer 05/17/2021
Supplement Date Received by Manufacturer07/08/2021
Initial Report FDA Received Date06/03/2021
Supplement Report FDA Received Date07/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-