• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA USA, INC. PROGRAMMING SOFTWARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA, INC. PROGRAMMING SOFTWARE Back to Search Results
Model Number 3000
Medical Device Problem Codes False Alarm (1013); Application Program Problem: Parameter Calculation Error (1449)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Voluntary medical device correction (remedial action) was initiated by the manufacturer on 2/4/2020, and the physician was provided a notification letter with recommended actions.Internal reference number: (b)(4).
 
Event or Problem Description
It was reported that the patient's device showed a current not delivered message after turning the device back on after mri.System diagnostics were normal, the error only occurred when an in-session re-interrogation was performed.The software version of the tablet has not yet been confirmed however this is likely related to a known software error.Internal testing and data review identified that false low output current messages can occur when m3000 (b)(4) software programs a m103-106 generator after a specific programming sequence occurs.When m3000 (b)(4) tablets initiate a programming session with a m103-106 generator, where its output current is programmed off (0 ma), and then the output current is programmed back on to >0 ma, a low output current message would be seen when an in-session interrogation is performed.This would persist upon multiple in-session interrogations and if diagnostics weren't performed during the session, show on the session report.Running system diagnostics or ending and restarting the session will confirm functionality of the device, and resolve the ¿output current low¿ error message.The low output current messages in these cases are false and do not impact generator function.Based on the reported programming sequence that occurred that day, it was consistent with the false low output current mechanism described above.Therefore, the patient's generator is expected to be able to deliver therapy as programmed.No further relevant information has been received to date.
 
Event or Problem Description
The software version was confirmed to be m3000 v1.6.No further relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Common Device Name
PROGRAMMING SOFTWARE
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
MDR Report Key11936362
Report Number1644487-2021-00764
Device Sequence Number4176001
Product Code LYJ
UDI-Device Identifier05425025750443
UDI-Public05425025750443
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P970003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3000
Was Device Available for Evaluation? No
Event Location Other
Initial Date Received by Manufacturer 05/11/2021
Supplement Date Received by Manufacturer06/17/2021
Initial Report FDA Received Date06/03/2021
Supplement Report FDA Received Date07/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age33 YR
-
-