• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/07/2021
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could not reproduce the reported issue.The potentially involved components, motor control board and motor end-stage board, were replaced.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s5 double head pump gave a motor control failure error message during bypass.The user tried turning off and back on the device as well as the s5 system but the issue persisted.The pump was replaced with another one and the procedure could be completed.There was no report of patient injury.
 
Manufacturer Narrative
The involved pump is installed since 2020 and according to the review of the complaint database no further events have been reported for this unit.The reported error message can be caused by: a defective motor controller board (hms 0409); a defective motor amplifier board (hmf 0408); a defective can communication between the hms an the hmf board.A livanova field service representative was dispatched to the customer's facility.The unit was checked on site according to maintenance procedures and no deviation was detected: the roller pump worked as expected and within specification.Thus, hardware malfunctions can be ruled out as cause of the event.Based on the above facts, it cannot be ruled out that the issue was caused by an intermittent communication between the hms and hmf board.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
MDR Report Key11937399
MDR Text Key268102220
Report Number9611109-2021-00303
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Was the Report Sent to FDA? No
Date Manufacturer Received08/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-