Model Number EP003994S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/06/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during a procedure, when performing the transseptal puncture, the transseptal needle did not progress easily through the dilator.It was noted the needle had transfixed the dilator.Two additional dilators were used without resolving the issue.It was noted that when the needle was seen at the distal tip of the dilator, plastic remnants were seen.The needle was replaced once, but the issue persisted.The needle was replaced again to resolve the issue.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: photos of a needle and dilator were received for analysis.The photo showed that the dilator was punctured by the needle, and confirmed broken plastic dilator pieces.In conclusion, the physical product was not returned.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROCKENBROUGH NEEDLE
Common Device Name
TROCAR
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
MDR Report Key11938863
Report Number9617601-2021-00109
Device Sequence Number1297187
Product Code DRC
UDI-Device Identifier00613994663146
UDI-Public00613994663146
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
PRE-AMEND
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/06/2021
Device Model NumberEP003994S
Device Catalogue NumberEP003994S
Device Lot Number218234907
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/06/2021
Supplement Date Received by Manufacturer09/06/2021
Initial Report FDA Received Date06/04/2021
Supplement Report FDA Received Date10/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EP003994S NEEDLE
-
-