• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, FRENCH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, FRENCH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-64
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/11/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate the unit and after checking the pump, the logs and having a conversation with the users about the issue, it appears that it was all a use problem.According to what the users claimed, they were surprised to see the pump stopped to perform the balloon fo calibration.Their first reaction began by touching "many keys", they finished with no augmentation (unit "not pumping") and the low diastolic pressure alarm.The users finally swapped to "manual" mode and the patient could be transferred with no injury.When the investigation to the user was completed, the fse complete a preventive maintenance (pm) on the machine.Unit passed all functional and safety test per factory specifications.The iabp was released to the customer and cleared for clinical service.A supplemental report will be submitted when additional information is provided to us.
 
Event Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) had a fiber optic problems and the user had to move it to make it work.The patient was transferred to another hospital with no injury linked to the unit.The service engineer will check this iabp during the normal pm.
 
Manufacturer Narrative
Updated fields; b4, b6, b7, d4, d9, g3, g6, g7, h2, h6, h10, h11.Corrected fields: d1, g1.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, FRENCH, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key11939434
MDR Text Key259720798
Report Number2249723-2021-01180
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108308
UDI-Public10607567108308
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-64
Device Catalogue Number0998-00-3023-64
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received01/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/27/2004
Is the Device Single Use? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SENSATION BALLOON
Patient SexPrefer Not To Disclose
-
-