| Model Number |
NIITP6510 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Inflammation (1932)
|
| Date of Event |
03/14/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet (b)(4).Patient weight: not provided.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that implant was removed due to peri-implantitis at tooth location 15.Inflammation was also reported.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint (b)(4).The following sections have been updated: d9: device available for evaluation change ¿no' to yes.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint cmp-(b)(4).A device history record review was performed and no related deviations or nonconformances were noted.Also, a complaint history search was performed using our complaint handling system and there were no additional related complaints for this product lot.The following sections have been updated: h6: evaluation codes h10: additional narrative.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Search Alerts/Recalls
|