COVIDIEN 20FR PEG ADAPTOR ENFIT; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS Back to Search Results
Model Number 8884752042E
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/03/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.
 
Event or Problem Description
The customer reported the y port adaptor leaks at the point where it connects to the funnel.No patient injury was reported.
 
Additional Manufacturer Narrative
Additional information: h6 type of investigation; investigation findings; investigation conclusions.Investigation summary: one decontaminated y-port without original package or lot number was received for evaluation.After performing the visual and functional evaluations, the reported condition was confirmed.Since a lot number was not provided, the device history record could not be performed.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.A review through the manufacturing process was conducted.All process and controls were found properly followed.There were no abnormal conditions found that could trigger the reported condition.The manufacturing process includes 100% visual inspection by the production personnel during the packaging process in order to detect and discard any identified failures.Based on the available information, an exact root cause could not be determined and no trend exists for the reported issue.At this time, no action plan is deemed required.Functional testing and visual inspections are being performed according to our current quality standards and inspection procedures.We will keep monitoring the process for any adverse trends that require immediate attention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
20FR PEG ADAPTOR ENFIT
Common Device Name
GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS
Manufacturer (Section D)
COVIDIEN
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX  22500
MDR Report Key11941866
Report Number9612030-2021-02928
Device Sequence Number4226984
Product Code PIF
UDI-Device Identifier30884521664061
UDI-Public30884521664061
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number8884752042E
Device Catalogue Number8884752042E
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 06/03/2021
Supplement Date Received by Manufacturer06/03/2021
Initial Report FDA Received Date06/04/2021
Supplement Report FDA Received Date09/02/2021
Patient Sequence Number1
-
-