MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Charging Problem (2892)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/13/2021
Event Type  malfunction  
Manufacturer Narrative
The full name of the event site was shortened due to field character limit; the full name is (b)(6) hospital.A getinge field service engineer (fse) done video call with the biomed, the fse was able to observe the battery was charging and unit did not have any battery issue while plugged in and while not plugged in.Therefor there was no issue with battery charging.Upon completion of our investigation, a supplemental report will be submitted.
 
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) would not charge while it was plugged in.It is unknown the circumstances under which the event occurred.There was no patient involvement; thus, no adverse event was reported.
 
Manufacturer Narrative
*the investigation type in the initial report was submitted incorrectly and 'communication/interviews|4111' should have been selected.As there was no issue with the battery as observed by fse on video call, the unit was cleared for clinical use and returned to the customer.Analysis of production: (3331/213) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 24 month product complaint trend data for the period jun 2019 through may 2021 was reviewed.There were no triggers identified for the review period.
 
Event Description
Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 
MDR Report Key11942338
MDR Text Key254872944
Report Number2249723-2021-01199
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/17/2008
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
NA.
-
-