| Model Number |
802414 |
| Medical Device Problem Codes |
Incorrect Measurement (1383); Material Protrusion/Extrusion (2979)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
05/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient stated the end of the bardia urethral catheter was too big and patient was bleeding sometimes.No medical intervention in event.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed manufacturing related.1 sample was confirmed to exhibit the reported failure.The device had not met specifications.The product was used for treatment purposes.A potential root cause for this failure could be "operator or machine error".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review is not required as the reported event was confirmed manufacturing related.Corrections: f h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was evaluated.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient stated the end of the bardia urethral catheter was too big and patient was bleeding sometimes.No medical intervention in event.
|
| |
|