• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® URETHRAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® URETHRAL CATHETER Back to Search Results
Model Number 802414
Medical Device Problem Codes Incorrect Measurement (1383); Material Protrusion/Extrusion (2979)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the patient stated the end of the bardia urethral catheter was too big and patient was bleeding sometimes.No medical intervention in event.
 
Additional Manufacturer Narrative
The reported event was confirmed manufacturing related.1 sample was confirmed to exhibit the reported failure.The device had not met specifications.The product was used for treatment purposes.A potential root cause for this failure could be "operator or machine error".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review is not required as the reported event was confirmed manufacturing related.Corrections: f h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was evaluated.
 
Event or Problem Description
It was reported that the patient stated the end of the bardia urethral catheter was too big and patient was bleeding sometimes.No medical intervention in event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDIA® URETHRAL CATHETER
Common Device Name
URETHRAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key11942435
Report Number1018233-2021-03253
Device Sequence Number4365288
Product Code KOD
UDI-Device Identifier00801741018350
UDI-Public(01)00801741018350
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/29/2024
Device Model Number802414
Device Catalogue Number802414
Device Lot NumberNGDV0114
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/13/2021
Supplement Date Received by Manufacturer11/19/2021
Initial Report FDA Received Date06/04/2021
Supplement Report FDA Received Date12/06/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/20/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-