| Model Number |
LS-TU7S5560 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Based on discussion on 04may2021 with the clinical specialist who was present at the initial procedure, the surgeon had attempted in-situ assembly of the tulips.The tulip associated with this failure, as well as the adjacent screw, was returned to integrity implants.Upon receipt and inspection of the device, including a comprehensive review of the device manufacturing history, it was confirmed that the device was manufactured to design specification.There is evidence to suggest that the surgeon was potentially impeded from mating the tulip and screw shank due to visual or anatomical constraints.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an initial posterior fixation procedure on (b)(6) 2021.Subsequently, post-operative images, obtained on (b)(6) 2021, revealed dissociation of the linesider standard tulip from the screw shank.A revision procedure was conducted on (b)(6) 2021.
|
| |
|