SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
Model Number 72200616 |
Device Problems
Overheating of Device (1437); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that during set up for an arthroscopy, the motor drive unit had a mechanical noise and was overheating.The procedure was successfully completed without delay using a smith and nephew back up device.No patient injury or other complications were reported.
|
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).The reported device, intended for use in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A relationship, if any, between the subject device and the reported event could not be determined.No containment or corrective actions are recommended at this time.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A complaint history review concluded this was a repeat issue.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
|
Manufacturer Narrative
|
H3, h6 the reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the exterior of product and the soak cap is missing and the cord is discolored.Mdu failed functional tests with blade stall error and overheating of cord.After troubleshooting, the cause of the overheating was observed to be a defective power cord assembly.It was determined that the power cord has shorted internal wiring.The complaint investigation has concluded that this unit has succumbed to physical damage to power cord.Factors which can contribute to a shorted power cord include rolling a heavy cart over the cord or excessive bending of the cord.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
|
|