|
The subject device was returned to olympus medical systems corp.(omsc) for investigation.Scratches were observed at the distal end of the probe.The surface of the probe was worn out severely, and the metal surface was exposed.Also, scratches were observed on the metal surface.The location of the scratches on the metal surface was matched with the position of the scratches on the distal end of the probe.Investigation was carried out by connecting aomori olympus sonosurg-g2, t2h-c, maj-1121, and maj-51 to the returned device to confirm the condition of the ultrasonic output activation.The output was activating without any problems, and the ultrasonic output warning did not occur.The probe holder was detached.Manufacturing record investigation: the dhr with the serial number of the subject device was confirmed.No abnormalities detected in the dhr for the following items which related to the reported phenomenon.Ultrasonic output.Damage of the probe during normal use.Actuation test - ultrasonic oscillation.Amplitude value test amplitude.Appearance test - appearance.Labeling review: the content of the instruction manual (drawing number: ge3691, revision number: 19 ) was confirmed.The instruction manual contains the necessary and enough information to handle the device as follows: do not activate output when closing the instrument and nothing is grasped between the grasping section and the probe, or when it is not possible to confirm that the tissue being grasped has been completely resected.Otherwise, abnormal heat generated by friction between the grasping section and the probe may damage the grasping section, or cause it to detach or premature wear in the grasping surface.Before each case, prepare and inspect this instrument as instructed below.Inspect other equipment to be used with this instrument as instructed in their respective instruction manuals.Should the slightest irregularity be observed, do not use this instrument and see chapter 8, troubleshooting.If the irregularity is still observed after consulting chapter 8, contact olympus.Damage or irregularity may compromise patient or user safety and may result in more severe equipment damage.Examine the instrument thoroughly after each use.If any abnormality is detected, do not use the instrument, and contact olympus.Conclusion summary: based on the investigation result, a likely factor of the event might be the following: the grasping surface was severely worn out.It can be inferred that the ultrasonic output was possibly activating while the metal area under the grasping surface was contacting the probe, causing scratches on the probe.It can be inferred that the ultrasonic output was repeatedly activated while the grasping section was closed without grasping anything in between the grasping section and the distal end of the probe, or the ultrasonic output was repeatedly activated when completion of the tissue dissection was not confirmed.This might have caused the grasping surface to wear out severely.A likely mechanism causing detachment of the probe holder might be the following: due to the heat of autoclave, the adhesive agent which affixes the pipe deteriorated.A device such as a brush with a thin tip was inserted from the distal end of the insertion portion for cleaning.This pressed the probe holder, causing the peeling.Due to the repeated used of the device with the condition stated 1 and 2, the peeling progressed.As a result, the probe holder detached.
|