• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, Back to Search Results
Model Number 71335552
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Extraskeletal ossification (1428); Pain (1994)
Date of Event 09/29/2014
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that, 6 moths after a revision surgery performed on (b)(6) 2013 (covered under (b)(4)), the patient reported increasing pain.The patient was treated with physical therapy and steroid injections, with no success.Radiographic imaging revealed that the components are well fixed and that the patient had heterotrophic ossification.The physician is concerned that the lower extremity pain is coming from the patient¿s back.No other complications were reported.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, the root cause for the heterotopic ossification cannot be concluded but some patients can be genetically predisposed it is a known complication of joint surgeries and is related to the procedure and not the device.The patient¿s lower back is the source of his pain and clinical symptoms per the report and is not associated with a malperformance of the implant.No further clinical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A complaint history review found related failures for the listed batch; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, joint tightness or patient condition.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 3 HOLE ACET SHELL 52MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED,
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11947641
Report Number1020279-2021-05029
Device Sequence Number19097725
Product Code MRA
UDI-Device Identifier03596010598240
UDI-Public03596010598240
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K070756
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/24/2021
Device Model Number71335552
Device Catalogue Number71335552
Device Lot Number11FM16286
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/14/2021
Supplement Date Received by Manufacturer09/03/2021
Initial Report FDA Received Date06/07/2021
Supplement Report FDA Received Date09/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
-
-