Model Number 24686
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 70% stenosed target lesion was located in the non-tortuous and moderately calcified brachial artery.A 12.0 x40, 75cm gladiator balloon catheter was advanced.However, during first inflation at 10 atmospheres for 20 seconds, the balloon ruptured.The procedure was completed with another of the same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: a gladiator 12 x 40, 14atm, 75cm was returned for analysis.The sheath used by the customer was not returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A partial balloon circumferential tear was identified located approximately 13mm distal of the proximal balloon markerband.From the partial circumferential tear, the balloon was also torn longitudinally for approximately 48mm proximally.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.A visual and microscopic examination observed no damage to the tip of the device.A visual and tactile examination found no kinks or damage to the shaft of the device.Both markerbands were undamaged and present on the shaft of the device.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 70% stenosed target lesion was located in the non-tortuous and moderately calcified brachial artery.A 12.0 x40, 75cm gladiator balloon catheter was advanced.However, during first inflation at 10 atmospheres for 20 seconds, the balloon ruptured.The procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11947745
Report Number2134265-2021-07327
Device Sequence Number1153010
Product Code LIT
UDI-Device Identifier08714729809555
UDI-Public08714729809555
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K132810
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/08/2024
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0026936015
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2021
Initial Date Received by Manufacturer 05/25/2021
Supplement Date Received by Manufacturer06/21/2021
Initial Report FDA Received Date06/07/2021
Supplement Report FDA Received Date07/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age83 YR
-
-