MAUDE Adverse Event Report: ARTHREX, INC. GLENOID, PE WITH PEG, SMALL; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. GLENOID, PE WITH PEG, SMALL; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number GLENOID, PE WITH PEG, SMALL
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/17/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that a revision surgery was necessary due to a aseptic loosening of the glenoid.The shoulder prosthesis was changed to an inverse shoulder prosthesis.No further information received.On 20-may-2021 update: further information were provided that the initial implantation was performed on (b)(6) 2016.During the revision it was found that the rotator cuff was intact and that there was no sign of an infection or other morphology of the bone.The patient had a strain active life style and work.The patient was treated with a tornier aequalis reversed ii with a glenoid structure.
 
Additional Manufacturer Narrative
The device was not returned.An x-ray was provided, however the x-ray did not provide enough information to confirm or investigate the complaint.
 
Additional Manufacturer Narrative
Based on the information provided, which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to a patient-specific event that may occur due to fatigue wear of the implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLENOID, PE WITH PEG, SMALL
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key11947845
Report Number1220246-2021-03232
Device Sequence Number4420735
Product Code KWS
UDI-Device Identifier00888867057319
UDI-Public00888867057319
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K083435
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2020
Device Model NumberGLENOID, PE WITH PEG, SMALL
Device Catalogue NumberAR-9105-01
Device Lot Number1516007
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2021
Supplement Date Received by Manufacturer05/19/2021
08/22/2024
Initial Report FDA Received Date06/07/2021
Supplement Report FDA Received Date12/22/2021
09/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/01/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-