Model Number 693575
Medical Device Problem Codes Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Low impedance (2285); Capturing Problem (2891); Impedance Problem (2950)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 05/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient experienced inappropriate therapy.The right ventricular (rv) lead triggered a lead integrity alert (lia) for high rate non-sustained episodes and a high number of short v-v intervals.In addition, the rv lead triggered alerts for oversensing of noise, high pacing impedance, and low pacing impedance.The rv lead also had an acute decrease in pacing impedance and thresholds were elevated.All ventricular fibrillation (vf)/ ventricular tachycardia (vt) detections were programmed off.The lead remains in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE S MRI SURESCAN
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key11947944
Report Number2649622-2021-11317
Device Sequence Number4491811
Product Code LWS
UDI-Device Identifier00613994446114
UDI-Public00613994446114
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/17/2016
Device Model Number693575
Device Catalogue Number693575
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/22/2021
Supplement Date Received by Manufacturer06/16/2021
Initial Report FDA Received Date06/07/2021
Supplement Report FDA Received Date06/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DVBB1D1 ICD; DVBB1D1 ICD
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient Age70 YR
Patient Weight139
-
-