COOPERSURGICAL, INC. LEEP SYSTEM 1000 ESU GEN Back to Search Results
Model Number 52969
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem Insufficient Information (4580)
Event Date 05/24/2021
Event Type  malfunction  
Manufacturer Narrative
Coopersurgical , inc.Is currently investigating the reported conditon.
 
Event Description
Malfunctioning diaphragm.Order: (b)(4).Leep system 1000 esu gen 52969.E-complaint-(b)(4).
 
Manufacturer Narrative
Investigation.X-inspect returned samples.*analysis and findings complaint #(b)(4).Distribution history: this complaint unit was manufactured at csi on 06/25/2010 under wo #50149 and shipped on 04/20/2006.Manufacturing record review: a review of the device history record could not be performed as the record could not be located at this time.However, at the time of manufacture, records from each unit are reviewed to ensure that product meet all specifications.Should the device history record be located, this complaint will be amended accordingly.Incoming inspection review: n/a service history record: no additional service history records found for this unit.Historical complaint review: a review of the 2-year complaint history showed similar reported complaint conditions.Product receipt: the complaint unit was returned on repair log 96384.Visual evaluation: visual examination of the complaint unit revealed no physical damage.Functional evaluation: unit was not functioning properly.Root cause: the diaphragm is the mechanism to make contact (for electrical current) to supply power to the unit.Air displacement pushes on a piston via the diaphragm into a switch to turn the power on.The foot pedal creates the air displacement to the pneumatic switch mechanism located in the housing of the unit.The diaphragm breaks down and loses its seal and cannot function properly.The diaphragm is described as a rubber, possibly a latex.Given the appearance of a bad diaphragm it appears to have been exposed to excessive stimuli.Materials like latex are flexible hence the use in this application, but its' elastic properties can alter over time as well.The root cause for this complaint condition is component related to the diaphragm.*correction and/or corrective action coopersurgical service and repair replaced the diaphragm on the unit, tested and returned it to the customer.Engineering has successfully tested a replacement material made of silicone for use in repairs going forward, eng-test-10341-r.The ifu was also updated to add a safety check via ecn-20444.A service bulletin was issued to existing customers informing them to check for this issue and return the unit if needed.All product in fg and sk, as applicable, were reworked to replace the previous versions of the dfus on all applicable products.The repaired unit will have been repaired with this new silicone material.
 
Event Description
Malfunctioning diaphragm.Order: (b)(4).1216677-2021-00125 leep system 1000 esu gen 52969 (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEEP SYSTEM 1000 ESU GEN
Type of Device
LEEP SYSTEM 1000 ESU GEN
Manufacturer (Section D)
COOPERSURGICAL, INC.
95 corporate drive
trumbull CT 06611
Manufacturer (Section G)
COOPERSURGICAL, INC.
75 corporate drive
trumbull CT 06611
Manufacturer Contact
michael marone
50 corporate drive
trumbull, CT 06611
4752651665
MDR Report Key11948241
MDR Text Key261678318
Report Number1216677-2021-00125
Device Sequence Number1
Product Code HGI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K952483
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial,Followup
Report Date 11/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number52969
Device Catalogue Number52969
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/25/2021
Initial Date FDA Received06/07/2021
Supplement Dates Manufacturer Received05/25/2021
Supplement Dates FDA Received11/04/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-